Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

Company
Fresenius Kabi Usa, Llc
Recall Initiated
November 3, 2025
Posted
December 1, 2025
Recall Number
Z-0848-2026
Quantity
15,862 units
Firm Location
North Andover, MA

Reason for Recall

Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.

Distribution

US Nationwide distribution in the sates of CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, UT, VA, WA, WI.

Lot / Code Info

Model Number: LVP-0004. UDI-DI: 00811505030320.

Root Cause

Labeling design

Action Taken

Fresenius Kabi notified consignees on about 11/03/2025 via emailed letter. Consignees were informed of the issue and instructed to account for additional backpressure if check valve is used when configuring the downstream occlusion pressure alarm setting, adjust the occlusion alarm pressure to allow for a more balanced approach between occlusion detection and avoidance of unnecessary alarms, and continue to follow all Instructions for Use and institutional policies regarding infusion pump setup, monitoring, and alarm response. Consignees were requested to ensure all relevant persons within the organization are informed about this letter and the actions as described per the enclosed Tip Sheet on How to Use a Check Valve with the Ivenix Pump & Administration Sets. Additionally, consignees were requested complete and return the Customer Reply Form. Fresenius Kabi plans to update the Instructions for Use to include the information presented in the "Required Actions for Users", "Important Note on Check Valve Usage" and/or Tip Sheet on "How to Use a Check Valve with the Ivenix Pump & Administration Sets" for which scheduling is still being finalized.

More recalls by Fresenius Kabi Usa, Llc

View all recalls by this company →