ADVIA Chemistry Microalbumin_2 (ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
August 15, 2024
Posted
September 18, 2024
Recall Number
Z-3170-2024
Quantity
19645 units
Firm Location
Tarrytown, NY

Reason for Recall

Atellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Mexico, Morocco, Netherlands, New¿Zealand, Norway, Oman, P.R.¿China, Pakistan, Paraguay, Peru, Poland, Portugal, Qatar, Republic¿Korea, Romania, Russian¿Fed., Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, United¿Kingdom, Uruguay, Vietnam, Zaire. ¿¿¿¿¿¿

Lot / Code Info

UDI-DI: 00630414517643 Lot Numbers: 612528, 616644, 620971, 623768, 636498, 640513, 658356, 668872

Root Cause

Nonconforming Material/Component

Action Taken

Siemens Healthcare issued Urgent Medical Device Corrections (UMDC) to US consignees ACHC24-05.A.US, ACHC24-05.A.US.CHC on 8/15/24 via FedX and Urgent Field Safety Notices (UFSN)to foreign consignees ACHC24-05.A.OUS, ACHC24-05.A.OUS.CHC to affected customers informing them of the issue and the impact to results. ADVIA Chemistry customers are advised that lot 665399 meets the IFU prozone claim. Customers will be instructed to discontinue use of and discard the impacted reagent lots indicated in the letter. Atellica CH customers will be advised that lots 242194, 242195, and 242321 meet the IFU prozone claim. Customers can continue to use the impacted products, with the understanding that patient samples with values above 9,500 mg/dL (95,000 mg/L) can result in falsely depressed results when using the impacted lots. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution Lot 665399 meets the IFU prozone effect claim. The manufacturing control system has been updated to ensure that there is no impact to future lots. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.