ADVIA 120/2120/2120i 3-in-1 TESTpoint Control (ABN1 Control)
- Recall Initiated
- June 4, 2025
- Posted
- July 3, 2025
- Recall Number
- Z-2087-2025
- Quantity
- 4,214 units
- Firm Location
- Tarrytown, NY
Reason for Recall
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Distribution
Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
Lot / Code Info
Siemens Material Number 10330063; UDI: 00630414473406, 00630414473406; Lot No. TP251035, TP251045.
Root Cause
Under Investigation by firm
Action Taken
An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.