RecallDepth

Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
August 11, 2025
Posted
September 8, 2025
Recall Number
Z-2542-2025
Quantity
11,811 units
Firm Location
Tarrytown, NY

Reason for Recall

Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Macedonia, Malaysia, Maldives, Mexico, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Uruguay, Vatikancity, Vietnam, Zaire.

Lot / Code Info

Material Number: 11097543. Test Code: UCFP. UDI-DI: 00630414279206. All lot numbers.

Root Cause

Under Investigation by firm

Action Taken

Siemens Healthcare Diagnostics notified consignees on about 08/11/2025 via letter. Customers were instructed to perform QC on each well, track additional reagent consumption (number of tests) to report to Siemens Healthineers for future reimbursement/credit, review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, and retain this letter with your laboratory records and forward this letter to those who may have received this product. Additionally, consignees were instructed to complete and return the Field Correction Effectiveness Check Form.

More recalls by Siemens Healthcare Diagnostics, Inc.

View all recalls by this company →