WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Company
Microvention Inc.
Recall Initiated
September 21, 2022
Posted
November 2, 2022
Recall Number
Z-0170-2023
Quantity
2018
Firm Location
Aliso Viejo, CA

Reason for Recall

Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels

Distribution

Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.

Lot / Code Info

REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345; Distributed in China: WDC/00842429113684/0000078840, 0000086571, 0000088347, 0000088349, 0000136477, 0000181460, 0000191147; Distribution in Japan: MV-WDC2/04987892122231/0000142172; 0000166445; 0000210109; 0000245337

Root Cause

Process control

Action Taken

On 09/21/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify, quarantine, and return all impacted devices. 2) Complete and return Customer Acknowledgment and Device Reconciliation form the recall instructions provided in the letter and have shared this notification with all device users within my facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred. Complete and return the acknowledgement and receipt form. Customers with additional questions can call the recalling firm at 949-501-0355, Monday - Friday PST 8:00 - 5:00 pm or email customerservice@microvention.com