BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70105.2794. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.

Company
Maquet Medical Systems, Usa
Recall Initiated
July 22, 2022
Recall Number
Z-1764-2022
Quantity
206 US; 48,960 Worldwide
Firm Location
Wayne, NJ

Reason for Recall

The firm has expanded the recall for the HLS Set Advanced regarding a potential breach in the sterility barrier that may result in compromised product sterility (prior recall was initiated November 6, 2020).

Distribution

Worldwide distribution - US Nationwide and the countries Canada and Australia.

Lot / Code Info

UDI-DI: 04037691741543 Lots 3000183172, 3000183179, 3000183191, 3000183910, 3000184681

Root Cause

Package design/selection

Action Taken

On July 15, 2022, the firm notified affected customers via Urgent Medical Device Removal Expansion letter. The letter informed customers of the inclusion of additional product codes and lots into the earlier recall initiated on November 6, 2020. Customers were instructed to examine their inventory and identify any affected product on hand. Affected product in use may be continued to be used according to normal practices. Unused and unexpired affected product should be immediately returned to Getinge for credit. Please contact Getinge Customer Service at (888) 9GETUSA / (888) 943-8872 (press option 2) between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. At this time, the recalling firm has already implemented the more durable Tyvek accessory packaging in all product manufactured September 2021 and later.

More recalls by Maquet Medical Systems, Usa

View all recalls by this company →