Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
This recall has been terminated (originally issued February 18, 2016).
- Company
- Medtronic
- Recall Initiated
- December 2, 2015
- Posted
- February 18, 2016
- Terminated
- July 18, 2019
- Recall Number
- Z-0821-2016
- Quantity
- 327,797 units
- Firm Location
- North Haven, CT
- Official Source
- View on FDA website ↗
Reason for Recall
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
Distribution
Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.
Lot / Code Info
Lot numbers within each range are affected: Lot # starting with P3C through P5B0454X, P5B0468X, P5C0152X through P5E0450X. Individual Lot Number: P5B0468X, P5E0523X, P5E0273X
Root Cause
Process control
Action Taken
Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.