RecallDepth

Sutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH POLYSORB 2/0 48" U/D DLU SU, Item Number: 170057; ENDO STITCH POLYSORB 2/0 48" VIO DLU SU, Item Number: 170053

This recall has been terminated (originally issued May 3, 2017).

Company
Medtronic
Recall Initiated
May 3, 2017
Terminated
July 18, 2019
Recall Number
Z-2543-2017
Quantity
478,465 (total for all products)
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

Potential for an incomplete seal on the inner package, which may result in premature degradation of the suture material.

Distribution

Worldwide distribution. US Nationwide including PR, and countries of AU, BE, CA, CL, CN, IL, IT, JP, KR, MX, PA, SA, SG, and TR.

Lot / Code Info

ENDO STITCH* POLYSORB* 0 48" U/D DLU SU J4M0853X J4M0855X J5A0192X ENDO STITCH* POLYSORB* 0 48" VIO DLU SU J4L1477X ENDO STITCH*POLYSORB* 2/0 48" U/D DLU SU J5A1097X ENDO STITCH*POLYSORB* 2/0 48" VIO DLU SU J5A0702X J5A0703X J5A0704X J5A0705X J5A1206X J5A1207X

Root Cause

Process control

Action Taken

All consignees were notified via Federal Express or certified mail on May 3, 2017, and the letter informs customers of the potential for an incomplete seal on the inner package and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice.

More recalls by Medtronic

View all recalls by this company →