Sutures. MAXON: MAXON 4/0 5X18" GREEN CV-24 D-TACH; Item Number: 8886621933; MAXON 5/0 5X18 GR CV-23 D-TACH; Item Number: XX2109; MAXON* 4/0 5X30 GREEN V-20 D*TACH; Item Number: 8886603433

This recall has been terminated (originally issued May 3, 2017).

Company
Medtronic
Recall Initiated
May 3, 2017
Terminated
July 18, 2019
Recall Number
Z-2542-2017
Quantity
478,465 (total for all products)
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

Potential for an incomplete seal on the inner package, which may result in premature degradation of the suture material.

Distribution

Worldwide distribution. US Nationwide including PR, and countries of AU, BE, CA, CL, CN, IL, IT, JP, KR, MX, PA, SA, SG, and TR.

Lot / Code Info

MAXON 4/0 5X18" GREEN CV-24 D-TACH D4K0396X D4K2053X D4L0541X D4L1084X D4L1272X D4M0394X MAXON 5/0 5X18 GR CV-23 D-TACH D4K1485X MAXON* 4/0 5X30 GREEN V-20 D*TACH D4L1156GX

Root Cause

Process control

Action Taken

All consignees were notified via Federal Express or certified mail on May 3, 2017, and the letter informs customers of the potential for an incomplete seal on the inner package and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice.