ASP-1020 LipoTower with Aspiration Pump. Medical use.
This recall has been terminated (originally issued September 30, 2015).
- Recall Initiated
- June 12, 2015
- Posted
- September 30, 2015
- Terminated
- May 23, 2016
- Recall Number
- Z-2821-2015
- Quantity
- 3 LipoTower with Aspiration Pump
- Firm Location
- Charlottesville, VA
- Official Source
- View on FDA website ↗
Reason for Recall
There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.
Distribution
Nationwide Distribution.
Lot / Code Info
Serial numbers 14031102, 15021105, and 14101103.
Root Cause
Device Design
Action Taken
The firm has issued FDA recall letters to all consignees. The letter have been sent out by UPS to expedite the delivery.