MicroAire K-Wires

This recall has been terminated (originally issued March 3, 2017).

Company
Microaire Surgical Instruments, Llc
Recall Initiated
February 21, 2017
Posted
March 3, 2017
Terminated
July 14, 2017
Recall Number
Z-1407-2017
Quantity
Total 362
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

Mislabeling on three lots of K-Wires. The incorrect product description " dual trocar" instead of "single trocar" was on the product label on the following parts: 1600-9355NS (lot 61005), 1600-9625NS (lot 60297), and 1604-162NS (lot 62856).

Distribution

United States.

Lot / Code Info

Part Number, Lot Number, Quantity Distributed to Field 1600-9355NS, 61005, 56; 1600-9625NS, 60297, 209; 1604-162NS, 62856, 97.

Root Cause

Labeling mix-ups

Action Taken

The firm will mail recall letters to the customers who received the mislabeled k-wires on February 21, 2017. Letters will be sent by UPS to expedite the delivery. The firm will contact the non-responding accounts at least two more times.