SST TWIST DRILL, STANDARD, 1.6mm Dia. x 127mm, REF 8054-010, STERILE This product line is intended for use in orthopaedic surgery (large and small bone surgeries) to cut bone

This recall has been terminated (originally issued May 22, 2017).

Company
Microaire Surgical Instruments, Llc
Recall Initiated
May 22, 2017
Terminated
August 29, 2018
Recall Number
Z-0412-2018
Quantity
42
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

One lot was found to contain both the correct twist drill (1.6mm) as well as an incorrect twist drill (2.0mm)

Distribution

US Distribution.

Lot / Code Info

LIT 0317305390

Root Cause

Labeling mix-ups

Action Taken

MicroAire Surgical Instruments, LLC distributed recall letters on May 22, 2017 MicroAire Surgical Instruments, LLC distributed recall letters to the customers who received the mislabeled twist. Letters will be sent by UPS to expedite the delivery. For further questions, please call (434) 975-8000