Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.

Company
Tornier, Inc
Recall Initiated
June 25, 2015
Posted
July 29, 2015
Terminated
August 31, 2017
Recall Number
Z-2315-2015
Quantity
99
Firm Location
Bloomington, MN

Reason for Recall

Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation will impact alignment and can affect the ability to fixate Aequalis IM Nail screws.

Distribution

US: AL, AK, AZ, CA, CO, CT, FL, IL, IN, ME, MD, MI, MO, NC, OH, PA, TX. OUS: CANADA AND FRANCE.

Lot / Code Info

Lots: DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102.

Root Cause

Nonconforming Material/Component

Action Taken

Consignees were sent on 6/25/2015 a Tornier "Urgent Medical Device Recall" letter dated June 24, 2015. The letter described the problem and the product affected by the recall. The letter provided information on jig inspection and referred consignee to the attached "Perfect Circles' surgical technique that should be used when separation is observed. Requested consignees to complete and return the questionnaire to FieldAction@tornier.com. For questions they can contact Customer Service Department at 1(888)494-7950.