stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder

Company
Tornier, Inc
Recall Initiated
August 29, 2024
Posted
September 16, 2024
Recall Number
Z-3155-2024
Quantity
5 units
Firm Location
Bloomington, MN

Reason for Recall

One lot of Perform Humeral Nucleus may exceed the specification for bacterial endotoxin load (max 20 EU/device).

Distribution

US Nationwide distribution in the states of IL, WI, AZ, NV, SD.

Lot / Code Info

UDI/DI 00846832092192, Lot number 3268BA

Root Cause

Under Investigation by firm

Action Taken

Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributor were directed to notify their customers. Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: identify and isolate all affected product to prevent accidental use, and complete and return the response form. Upon receipt, Stryker will contact the consignee and arrange for the return of the product. Distributors were directed to notify their customers. For questions or concerns, please contact memphis.fieldaction@stryker.com.