TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

Company
Tornier, Inc
Recall Initiated
December 19, 2025
Posted
February 2, 2026
Recall Number
Z-1237-2026
Quantity
427
Firm Location
Bloomington, MN

Reason for Recall

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

Distribution

US Nationwide distribution.

Lot / Code Info

Lot AZ4824290

Root Cause

Labeling mix-ups

Action Taken

On December 19, 2025, the firm began notifying affected customers via Urgent Field Safety Notice letters. Customers were instructed to immediately check their internal inventory for affected product. Firm stated that return of product is not necessary because the issue is detectable pre-operatively. The physical product itself is correctly marked, and after unpacking, should only be assigned to the correct place in the screw caddy. The notification is being provided for awareness only.