Segmented Cervix Applicator Set, part number GM11004310, a component used with the tube for segments and cervical sleeves The guide tube for segments and cervical sleeves in the Segmented Cervix Applicator Set was developed to treat cancer of the cervix, vagina, endometrium, vaginal stump and uterus. It is used for Brachytherapy.

This recall has been terminated (originally issued December 20, 2012).

Company
Varian Medical Systems, Inc.
Recall Initiated
September 28, 2012
Posted
December 20, 2012
Terminated
April 2, 2013
Recall Number
Z-0607-2013
Quantity
15 units
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

The guide tube which is part of the Segmented Cervix Applicator Set is difficult to disconnect from cervical sleeve.

Distribution

Worldwide Distribution - U.S. state of Georgia and the countries of Canada, Germany, India, Poland, and Thailand.

Lot / Code Info

Lot L33, Serial numbers H64Y009-GM, H64A056, H640576, H600272, H640784, H640774, H640764, H600603, H640547.

Root Cause

Device Design

Action Taken

The firm, Varian Medical Systems, sent an "URGENT MEDICAL DEVICE CORRECTION - URGENT FIELD SAFETY NOTICE" letter dated September 28, 2012 to all affected consignees/Customers. The letter identified the affected product, description of problem and recommended user actions to be taken. The consignees/customers were instructed to immediately remove all affected guiding tubes from clinical use, and sequester the components for pending return to Varian Brachy Therapy. Consignees/Customers were requested to complete the attached proof of notification form and return it to Varian Medical Systems. For questions contact-AT&T international direct dialing network (http://www.business.att.com/bt/dial_guide.isp. Then after hearing the AT&T signature noise, and. "Please enter the number you are calling", dial 800-360-7909; UK:+44 1293-601-327 or US:+1-434-979-1540. To submit a support request or general inquiry, compose an email with a blank subject line to: brachyhelp@varian.com.