VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA

This recall has been terminated (originally issued October 24, 2013).

Company
Varian Medical Systems, Inc.
Recall Initiated
August 9, 2013
Posted
October 24, 2013
Terminated
April 10, 2014
Recall Number
Z-0065-2014
Quantity
3 sets
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.

Distribution

Worldwide distribution: US (nationwide) distribution in CA and in the country of India.

Lot / Code Info

Lots #6078184-1, #6078184-2, #6078184-4

Root Cause

Nonconforming Material/Component

Action Taken

Urgent Medical Device Correction, Urgent Field Safety Notice were mailed by FedEx tracked delivery to all consignees on August 12, 2013.