Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.

This recall has been terminated (originally issued June 23, 2015).

Company
Varian Medical Systems, Inc.
Recall Initiated
May 18, 2015
Posted
June 23, 2015
Terminated
January 26, 2016
Recall Number
Z-1829-2015
Quantity
46
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat

Distribution

Worldwide Distribution.

Lot / Code Info

JAQ

Root Cause

Process design

Action Taken

All the consignees affected by this recall will be notified by: The Urgent Medical Device Correction Letter/ Field Safety Notification will be distributed to all affected users, with a description of the problem. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. Additionally, the letter will be posted to the customer support site: http:/ /www.MyVarian.com.

More recalls by Varian Medical Systems, Inc.