CELL-DYN Shear Valve Assembly used on CELL-DYN 3700CS Analyzer and CELL-DYN 3700CS Analyzer (refurb). Abbott Diagnostics Division, Santa Clara, CA 05054 Designed for in vitro diagnostic use in clinical laboratories.

Company
Abbott Laboratories
Recall Initiated
March 15, 2011
Posted
April 29, 2011
Terminated
November 22, 2011
Recall Number
Z-2014-2011
Quantity
174 total instruments, all varieties.
Firm Location
Santa Clara, CA

Reason for Recall

The Nylon Washer has an outer diameter which is undersized resulting in the inability to generate patient results.

Distribution

Worldwide Distribution -- AZ, CA, FL, IA, ID, IN, MA, MI, MN, MS, ND, NY, OK, OR, SC, SD, TX, WI, WY and Germany, Singapore, Mexico, China, South Korea, Argentina, Japan, Brazil, Canada, Australia, and Chile.

Lot / Code Info

List Number 02H30-01; lot number 20071AN; Refurbished list number: 02H030-03; lot numbers 20478AN96, 20390AN96, and 20421AN96.

Root Cause

Pending

Action Taken

Product Correction letters were sent on March 15, 2011. via Federal Express. The letter identified the affected product, explained the problem, the patient impact and the necessary action that were to be taken. Customers were to review the letter and inform their Laboratory Manager of the pending action.; complete and return the Customer Reply Form; and save a copy of the letter for their records. If there were any questions, customers were to contact Customer Support at 1-877-4ABBOTT or their local hematology customer support representative for customers outside of the US.