XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.

This recall has been terminated (originally issued June 16, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
April 8, 2008
Posted
June 16, 2011
Terminated
April 8, 2014
Recall Number
Z-2577-2011
Quantity
75
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct.

Distribution

Worldwide distribution: USA (nationwide) and countries of: China, France, Germany, Japan, Mongolia, Netherlands, Nicaragua, Poland, Russia, South Korea, Switzerland, and Turkey.

Lot / Code Info

XiO Release 4.0.0 through 4.40.00

Root Cause

Software design

Action Taken

The firm, CMS, Inc., sent a "XiO Customer Advisory" letter dated April 8, 2008 to all affected customers. The letter describes the product, the problem, and the actions to be taken ( workaround). The letter also states the problem will be resolved in a future release. No release number or date was given. If you have any questions, please call 408-380-8023.