CELL-DYN Sapphire Hematology Analyzer, List numbers 08H00-01, 08H00-03 A multi- parameter, automated hematology analyzer designed for in vitro diagnostic use in counting and characterizing blood cells.

Company
Abbott Laboratories
Recall Initiated
August 25, 2008
Posted
November 21, 2008
Terminated
January 6, 2011
Recall Number
Z-0306-2009
Quantity
958 units
Firm Location
Santa Clara, CA

Reason for Recall

Exposed wire resulted in minor electrical shock and burn to the service technician

Distribution

Worldwide Distribution --- including USA and countries of Argentina, Australia, Brazil, Canada, Chile, China WFOTC, Colombia, Germany, Hong Kong, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan and Venezuela.

Lot / Code Info

All pre production units with Serial numbers S-001 to S-026 and all production units with serial numbers 42003AZ-42007AZ, 42010AZ-42030AZ, 42032AZ-42033AZ, 42037AZ-42059AZ, 42061AZ-42062AZ, 42064AZ-42067AZ, 42072AZ-42074AZ, 42076AZ-42081AZ, 42083AZ-42123AZ, 42125AZ-42852AZ, 42854AZ-42944AZ, 42946AZ-42949AZ, 42951AZ, 42953AZ, 42955AZ-42961AZ, and 42963AZ-42964AZ

Root Cause

Device Design

Action Taken

Recall issued August 22, 2008. Technical service bulletin was issued for all field personnel via internal database. No customer communication is planned as customers are instructed *not* to remove the cover, which is necessary to access the area.