Medtronic Adjustable Valve, 6248VAL, Sterilized using irradiation. Medtronic, Inc. Minneapolis, MN 55432, USA. The Medtronic model 6248VAL adjustable valve is an accessory designed for use with Medtronic delivery systems to reduce blood loss during percutaneous catheter procedures. It consists of an adjustable valve and flush port.

This recall has been terminated (originally issued September 30, 2008).

Company
Medtronic Inc. Cardiac Rhythm Managment
Recall Initiated
August 27, 2008
Posted
September 30, 2008
Terminated
May 7, 2011
Recall Number
Z-2471-2008
Quantity
35,193 total
Firm Location
Mounds View, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic has identified an issue with the packaging for specific lots of the Medtronic Adjustable Valve and valves contained in the Attain Left Heart Delivery System kits. A small number of valve package seals could be compromised.

Distribution

Worldwide Distribution --- including USA and except states OR, RI, and SD. Puerto Rico and Washington D.C. , and countries of Algeria , Argentina , Australia , Austria , Belgium , Bosnia And Herzegovina , Canada , Chile , Croatia , Czech Republic , Denmark , Egypt , Finland , France , Georgia , Germany , Greece , Guadeloupe , Hong Kong , Hungary , India , Ireland , Israel , Italy , Japan , Jordan , Kazakhstan , Kuwait , Lithuania , Malta , Martinique , Morocco , Netherlands , Norway , Pakistan , Poland , Portugal , Puerto Rico ,Qatar , Reunion , Romania , Saudi Arabia , Slovakia , Slovenia , South Africa , Spain , Sri Lanka , Sweden , Switzerland , Syrian Arab Republic , Tunisia , Turkey , UK , Ukraine , United Arab Emirates , Uruguay , and Venezuela.

Lot / Code Info

Lot numbers: 33901607, 33907906, 33909207, 33913206, 33914207, 33917207, 33921308, 33925707, 33929306, 33931107, 33933205, 33934905, 33936105

Root Cause

Process control

Action Taken

An Urgent Device Recall letter, dated August 28, 2008, was sent to affected consignees. The letter describes the issue, identifies affected product, and requests that affected product be removed from inventory and returned to Medtronic. Medtronic representatives with assist in returning unused valves and kits and provide replacements. Contact Medtronic at 1-763-526-0000 for assistance.