Medtronic Virtuoso VR, Model D154VWC, Single Chamber implantable cardioverter defibrillator with atrial therapies. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. This device is indicated to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias in patients with NYHA functional Class II/III heart failure.
This recall has been terminated (originally issued August 18, 2009).
- Recall Initiated
- July 27, 2009
- Posted
- August 18, 2009
- Terminated
- May 31, 2012
- Recall Number
- Z-1954-2009
- Quantity
- 1
- Firm Location
- Mounds View, MN
- Official Source
- View on FDA website ↗
Reason for Recall
Medtronic has detected a specific pattern of MOSFET IC malfunctions in its Concerto, Virtuoso and EnRhythm family of devices. The probability of occurrence decreases with time and, to date over 90% of the malfunctions related to the pattern have occurred within the first twelve months after implant.
Distribution
Nationwide Distribution-- (states of IA and NJ).
Lot / Code Info
Serial number: PUN421618H.
Root Cause
Process control
Action Taken
Medtronic, Inc. personnel retrieved 2 non-implanted devices remaining at 2 hospitals. A Performance Notes letter was left at each facility. The Performance Notes described the issue, actions to address issue and recommendations to physicians. The Performance Notes will also be posted on the firm's website. Direct questions about this recall to Medtronic, Inc. by calling 1-763-514-4000.