RecallDepth

Medtronic Virtuoso DR, Model D154AWG, Dual Chamber implantable cardioverter defibrillator with atrial and ventricular therapies. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. This device is indicated to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias in patients with NYHA functional Class II/III heart failure. In addition, the device is indicated for use in the above patients with atrial tachyarrhythmias, or those patients who are at significant risk of developing atrial tachyarrhythmias.

This recall has been terminated (originally issued August 18, 2009).

Company
Medtronic Inc. Cardiac Rhythm Managment
Recall Initiated
July 27, 2009
Posted
August 18, 2009
Terminated
May 31, 2012
Recall Number
Z-1953-2009
Quantity
1
Firm Location
Mounds View, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic has detected a specific pattern of MOSFET IC malfunctions in its Concerto, Virtuoso and EnRhythm family of devices. The probability of occurrence decreases with time and, to date over 90% of the malfunctions related to the pattern have occurred within the first twelve months after implant.

Distribution

Nationwide Distribution-- (states of IA and NJ).

Lot / Code Info

Serial number: PUL450634H.

Root Cause

Process control

Action Taken

Medtronic, Inc. personnel retrieved 2 non-implanted devices remaining at 2 hospitals. A Performance Notes letter was left at each facility. The Performance Notes described the issue, actions to address issue and recommendations to physicians. The Performance Notes will also be posted on the firm's website. Direct questions about this recall to Medtronic, Inc. by calling 1-763-514-4000.

More recalls by Medtronic Inc. Cardiac Rhythm Managment

View all recalls by this company →