SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for surgical, GYN, NON-GYN and Autopsy cases.

This recall has been terminated (originally issued September 22, 2008).

Company
Scc Soft Computer
Recall Initiated
February 23, 2006
Posted
September 22, 2008
Terminated
November 20, 2008
Recall Number
Z-2329-2008
Quantity
335
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

Loss of text misrepresented individual tissue diagnosis on the final diagnosis print out from the HIS system software. The specimen heading "J" was removed and the diagnosis appeared under the specimen heading "I" as a result the specimen with the heading "I" had two diagnosis.

Distribution

Worldwide Distribution including USA and the country of Canada.

Lot / Code Info

Versions 3.17 and 4.1

Root Cause

Other

Action Taken

Software corrections were sent as of 10/30/2003 for clients to upgrade to, the correction software patch is in versions 3.17.5.16, 3.17.6.10, 3.17.7.3 and 4.1.1.0. The firm did not inform the non-affected consignees of the failure because the firm concluded that the associated risk was low.