Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light Green SF Yellowish Stain 225mL, Sodium Carbonate Solution 225mL, and Periodic Acid Solution 225mL. Each kit provides reagents sufficient for a maximum of 100 tests.
This recall has been terminated (originally issued December 27, 2006).
- Company
- Emd Chemicals Inc
- Recall Initiated
- July 25, 2006
- Posted
- December 27, 2006
- Terminated
- October 6, 2008
- Recall Number
- Z-0305-2007
- Quantity
- 74 kits
- Firm Location
- Gibbstown, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Due to loose caps, the SO2 levels dropped significantly from release levels causing the product not to perform properly.
Distribution
Nationwide.
Lot / Code Info
Item Number 64945/93 Lot # 5341, Exp. Date: May 31, 2007.
Root Cause
Other
Action Taken
On July 27, 2006 EMD contacted its direct account customers via letter. The firm requested discontinue use of the product and contact EMD for return.