LifeScan brand OneTouch Ultra Meter - Owner''s Booklet; Product is manufactured and distributed by LifeScan, Inc., 1000 Gibraltar Dr., Milpitas, CA 95037

Company
Lifescan Inc
Recall Initiated
November 21, 2003
Posted
September 29, 2005
Terminated
November 29, 2005
Recall Number
Z-1596-05
Quantity
Approximately 572,000 units
Firm Location
Milpitas, CA

Reason for Recall

LifeScan brand OneTouch Ultra Meter - Owner's Booklets are missing instructions on how to set the Units of Measurement (UOM)--mg/dL versus mmol/L. This error has the potential to lead to acute hyperglycemic complications if the user fails to realize that 18mg/dL of glucose is equal to 1 (one) mmol/L.

Distribution

The recall product was distributed in the U.S. only. LifeScan manufactured and distributed OneTouch Ultra System Kits with the affected OneTouch Ultra Meter Owner''s Booklets between March 31, 2003 and October 3, 2003.

Lot / Code Info

Includes all System Kits shipped during June 2003 until December 2003

Root Cause

Other

Action Taken

Firm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner''s booklet.