IMPAX 6.2.1 Network Gateway Server, License Key. Catalog # ABC code: EXEGE000. The product is used to transfer medical imaging data. AGFA Corp., Greenville, South Carolina 2960.

Company
Agfa Corp.
Recall Initiated
December 11, 2007
Posted
July 1, 2008
Terminated
July 21, 2010
Recall Number
Z-1306-2008
Quantity
12 units
Firm Location
Greenville, SC

Reason for Recall

Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare.

Distribution

Nationwide Distribution - USA, including states of CA, GA, IA and SD.

Lot / Code Info

ABC Code EXEGE000

Root Cause

Release of Material/Component prior to receiving test results

Action Taken

On 12/11/07, AGFA Healthcare sent out an e-mail to all affected consignees informing them of the situation. The firm also arranged to install the corrected License Key on all affected units of the product.