IMPAX 6.2.1 Network Gateway Server, License Key. Catalog # ABC code: EXEGE000. The product is used to transfer medical imaging data. AGFA Corp., Greenville, South Carolina 2960.
- Company
- Agfa Corp.
- Recall Initiated
- December 11, 2007
- Posted
- July 1, 2008
- Terminated
- July 21, 2010
- Recall Number
- Z-1306-2008
- Quantity
- 12 units
- Firm Location
- Greenville, SC
Reason for Recall
Failure of Data Transmission: No imaging data could be transmitted by the device, because a key value needed for data transmission was incorrectly entered in the affected units by AGFA Healthcare.
Distribution
Nationwide Distribution - USA, including states of CA, GA, IA and SD.
Lot / Code Info
ABC Code EXEGE000
Root Cause
Release of Material/Component prior to receiving test results
Action Taken
On 12/11/07, AGFA Healthcare sent out an e-mail to all affected consignees informing them of the situation. The firm also arranged to install the corrected License Key on all affected units of the product.