RecallDepth

Jet Medical 4F x 60cm Single Lumen Polyurethane PICC Nursing Tray. Catalog number JDMPN4S-T

Company
Medical Components, Inc Dba Medcomp
Recall Initiated
September 15, 2005
Posted
October 28, 2005
Terminated
November 22, 2005
Recall Number
Z-0103-06
Quantity
10 units
Firm Location
Harleysville, PA

Reason for Recall

The product is labelled as containing a 4F catheter when the trays/kits actually contain a 5F catheter. The catheters will not fit into the sheath-dilator that was included in the tray. mislabeled as 4F

Distribution

The products were shipped to distributors in OH and MO.

Lot / Code Info

Lot number MABY360

Root Cause

Other

Action Taken

On 9/15/05, the recalling firm telephoned the two distributors to inform them of the problem and the need to return the product.