Architect Progesterone Reagent Kit; list 6C26-27 (100 tests) and 6C26-22 (400 tests); Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064

This recall has been terminated (originally issued April 24, 2003).

Company
Abbott Laboratories Hpd/Add
Recall Initiated
March 31, 2003
Posted
April 24, 2003
Terminated
August 19, 2003
Recall Number
Z-0782-03
Quantity
368 kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

Lot gives lower specimen results compared to results generated with other lots.

Distribution

Nationwide, Canada, Columbia, England, Japan, Australia, New Zealand and Singapore

Lot / Code Info

list 6C26-27, lot 95651M201 list 6C26-22, lots 95651M200, 95651M202

Root Cause

Other

Action Taken

Recalled by telephone calls and follow-up letters dated 3/31/03. The accounts were informed of the depressed patient specimen results, and were requested to discontinue use and destroy any of the three affected lots found in inventory. The accounts were advised to notify physicians of the lower patient results observed with these lots and suggest review of the patient results to determine any impact on patient management.