IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.

Company
Fresenius Kabi Usa, Llc
Recall Initiated
May 12, 2025
Posted
June 6, 2025
Recall Number
Z-1869-2025
Quantity
14,280 units (724 cases)
Firm Location
North Andover, MA

Reason for Recall

Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.

Distribution

US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.

Lot / Code Info

Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015.

Root Cause

Under Investigation by firm

Action Taken

Consignees received an "URGENT - VOLUNTARY RECALL" notification via email dated 5/11/25. This notification instructs consignees to discontinue use and distribution of affected devices immediately. Consignees are to check their inventory for affected devices and place them in quarantine until product can be returned for replacement. The provided recall notification is to be distributed to users of the affected devices. Consignees are instructed to return the completed Customer Reply Form via fax or email to facilitate return of affected devices for replacement. Consignees with any questions can contact Fresenius Kabi by phone at 855-354-6387x1 or by email at Ivenix_support@fresenius-kabi.com.