IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
February 22, 2022
Recall Number
Z-1015-2022
Quantity
10,755 boxes (533 US, 10,222 OUS)
Firm Location
Tarrytown, NY

Reason for Recall

Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.

Lot / Code Info

8 production lots: 463, 464, 466, 467, 469, 470, 471, 472, [UDI for 200 tests: (01)00630414961132(10)463(17)20220228, (01)00630414961132(10)464(17)20220331, (01)00630414961132(10)466(17)20220331, (01)00630414961132(10)467(17)20220531, (01)00630414961132(10)469(17)20220731, (01)00630414961132(10)470(17)20220930, (01)00630414961132(10)471(17)20221130, (01)00630414961132(10)472(17)20230228] [UDI for 600 tests: (01)00630414961149(10)463(17)20220228, 01)00630414961149(10)464(17)20220331, (01)00630414961149(10)466(17)20220331, (01)00630414961149(10)467(17)20220531, (01)00630414961149(10)469(17)20220731, (01)00630414961149(10)470(17)20220930, (01)00630414961149(10)471(17)20221130, (01)00630414961149(10)472(17)20230228]

Root Cause

Under Investigation by firm

Action Taken

Urgent Medical Device Correction was issued February 02, 2022 via Fed Ex. Notification provides instructions to customers to mitigate this sample carryover. Customers are requested to complete and return the Effectiveness Check questionnaire. Contact your local Siemens Healthineers technical support representative with any questions.