ARIES Flu A/B & RSV Assay , REF 50-10020, UDI # 00840487100158
- Company
- Luminex Corporation
- Recall Initiated
- February 20, 2019
- Posted
- August 15, 2019
- Recall Number
- Z-2284-2019
- Firm Location
- Austin, TX
Reason for Recall
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Distribution
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Lot / Code Info
Lot numbers and discontinue use by dates for Part Number 50-10020: AA7127 - Immediately; AA7206 - Immediately; AA7312 - 3/5/2019; AA7549 - 3/30/2019; AA7635 - 4/10/2019
Root Cause
Process control
Action Taken
On February 20, 2019, the firm sent a "Urgent: Medical Device Recall/Field Safety Notice" via email to all consignees. The customer notification informed customers of the following: 1. As an immediate action, Luminex is requesting that customers discontinue use of and discard the products listedin the recall notice immediately or by the Discontinue Use By date. Contact Luminex Global Support Services to obtain replacement material. Material that is still within its Discontinue Use By date may be used at the customers discretion. 2. Complete the Acknowledgment and Receipt form. You must complete and return this form even if you do not have any remaining product. Luminex Technical Support can assist you in completing this form. This information is essential in order to maintain recall effectiveness information required by US FDA. 3. Although no adverse events have been reported, adverse reaction or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 4. Please call the Luminex Global Support Services Team if you have any questions or concerns. Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) support@luminexcorp.com CAN-0233