RecallDepth

ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

Company
Luminex Corporation
Recall Initiated
November 17, 2021
Terminated
July 8, 2024
Recall Number
Z-0553-2022
Quantity
329
Firm Location
Austin, TX

Reason for Recall

Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.

Distribution

Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.

Lot / Code Info

Lot Numbers AB3600A and AB3560A

Root Cause

Nonconforming Material/Component

Action Taken

On 11/17/2021, the firm notified affected customers via letter titled, "URGENT: MEDICAL DEVICE RECALL ARIES HSV 1&2 Assay." The letter indicated, if you observe any leaks, please contact Luminex Global Support Services at support@luminexcorp.com. Please inform all those who need to be aware within your organization of this potential product performance issue. Please contact Luminex Global Support Services at support@luminexcorp.com at your earliest convenience if you have a Kit with an affected lot to arrange for its replacement, and to confirm the disposal of any remaining Kits in your inventory. We ask that you complete and return the attached acknowledgement letter so that we can confirm the destruction of recalled lots. Luminex Global Support Services, 1-877-785-2323 (U.S. and Canada}, +1-512-381-4397 (Outside U.S. and Canada), support@luminexcorp.com

More recalls by Luminex Corporation

View all recalls by this company →