RecallDepth

2.0/2.4 Cannulated Screw Countersink Part # 210-24-002 NON-STERILE and 3.0/4.0 Cannulated Screw Countersink Part # 210-40-002 NON-STERILE Product Usage: The Tiger Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis, and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.

This recall has been terminated (originally issued August 19, 2015).

Company
Trilliant Surgical Ltd.
Recall Initiated
July 20, 2015
Posted
August 19, 2015
Terminated
January 26, 2017
Recall Number
Z-2421-2015
Quantity
102 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

The cutting efficiency of the Countersinks was out of specification.

Distribution

US Nationwide Distribution in the states of: NY, CT, TN, TX, OH, IN, LA, NC, MO, MN, GA, FL, SC, WA, WI, IA, PA, CA, ND, OK, NJ, MI and MD.

Lot / Code Info

Lot numbers 1362-01 and 1364-01

Root Cause

Nonconforming Material/Component

Action Taken

Trilliant sent an Important Medical Device Advisory Notice letter dated February 15, 2013 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers of the option to exchange affected instruments at no charge. To initiate the exchange customers were instructed to contact your local Trilliant Surgical Sales Representative, contact the Customer Service Department directly at 1-800-495-2919, or via email at orders@trilliantsurgical.com.

More recalls by Trilliant Surgical Ltd.

View all recalls by this company →