Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
February 6, 2015
Posted
February 13, 2015
Terminated
October 17, 2016
Recall Number
Z-1116-2015
Quantity
103 units
Firm Location
Ann Arbor, MI

Reason for Recall

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Distribution

US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.

Lot / Code Info

HX2 Temperature Management System, 120V  Catalog # 809810  serial numbers: 00016, 0001 1116  HX2 Temperature Management System, 100V Catalog # 812231  serial numbers: 0001, 1002, PT0003, PT0004   HX2 Temperature Management System, 240V Catalog # 812240  serial numbers: 0001, 0002  HX2 Temperature Management System, Trial Kit Catalog # 809810TK  serial numbers: 1001 1042

Root Cause

Device Design

Action Taken

On 2/6/2015, URGENT MEDICAL DEVICE RECALL notifications were mailed to the affected hospitals. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call customer service at 1-800-521-2818.