POUCH/CARTON LABEL: ***REF 5023-138***9.5 x 25.5 x 0.4mm***QTY 1***MICRO SAGITTAL BLADE, FINE***STERILE***Rx Only***CONMED LINVATEC, CONMED LINVATEC 11211 CONCEPT BLVD LARGO, FL 33773-4908 USA***Made in USA*** To cut soft tissue or bone

This recall has been terminated (originally issued April 4, 2011).

Company
Linvatec Corp. Dba Conmed Linvatec
Recall Initiated
November 30, 2010
Posted
April 4, 2011
Terminated
November 9, 2011
Recall Number
Z-1880-2011
Quantity
415
Firm Location
Largo, FL
Official Source
View on FDA website ↗

Reason for Recall

On Nov 30, 2010 ConMed Linvatec, Largo, FL recalled their Micro Sagittal Blade, Fine Lot number 180209, Product Number 5023-138. Some units of Micro Sagittal Blade, Fine from Lot 180209 may have compromised seals in the sterile packaging.

Distribution

Worldwide Distribution - USA including CT, FL, GA, HI, IN, MO, MS, MT, NJ, NY, OH, PA, SC, and TX and the countries of Canada, and South Africa.

Lot / Code Info

Lot: 180209

Root Cause

Process control

Action Taken

Linvatec Corp. dba ConMed sent an Urgent Medical Device Recall Notification letter dated November 30, 2010, by FedEx, email, Fax, etc) to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete the reply form within 30 days of receipt. Immediately check their facility's inventory for the affected product and lot number. If they had any of the product with lot number 180209. Do Not Use. Segregate and return for credit. If the product was transferred to another facility, customers were to complete the REPLY FORM documenting the transfer of the product and fax the form to 727-319-5701. For questions regarding this recall call 727-319-5701 or 800-535-8536.