Ref: VP1600 HD Still Capture System ConMed Linvatec . Captures, stores and transfers images generated by a medical endoscopic camera system during surgical procedures.

This recall has been terminated (originally issued May 20, 2013).

Company
Linvatec Corp. Dba Conmed Linvatec
Recall Initiated
March 18, 2013
Posted
May 20, 2013
Terminated
November 6, 2013
Recall Number
Z-1340-2013
Quantity
267 units
Firm Location
Largo, FL
Official Source
View on FDA website ↗

Reason for Recall

The VP1600 HD Still Capture Image System is recalled due to an anomaly between the software code in the VP1600 console and the communications interface for various keyboard key strokes (Caps lock and Num Lock) which may lock up and requires the user to reboot to continue.

Distribution

Worldwide Distribution-USA (nationwide) including the Virgin Islands and the states of CA, CT, FL, GA, IN, KY, KS, LA, MD, MI, MN, MO, MS, MT, NE, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, and WI, and the countries of Austria, Belgium, Canada, Denmark, Hong Kong and Zambia.

Lot / Code Info

VP1600

Root Cause

Software design (manufacturing process)

Action Taken

The firm, ConMed, sent an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" letter dated 3/8/2013 to its customers after initial phone notifications. The letter identified the product, problem, and actions to be taken by the customers.