D3000I Advantage Drive System, 230v, surgical instrument motor. Product Usage: The Advantage Drive Systems function as a powered instrument system consisting of handpieces and accessories to perform cutting of soft tissue and bone

This recall has been terminated (originally issued March 6, 2013).

Company
Linvatec Corp. Dba Conmed Linvatec
Recall Initiated
January 24, 2013
Posted
March 6, 2013
Terminated
July 5, 2013
Recall Number
Z-0917-2013
Quantity
23 units
Firm Location
Largo, FL
Official Source
View on FDA website ↗

Reason for Recall

The Membrane Switch Panel (referred to as "front panel" of the component part number A34-003-207, used on both the D3000 and D3000I Advantage Drive System, contained the incorrect symbology per third edition IEC60606-01 Medical Electrical Equipment: General requirements for basic safety and essential performance.

Distribution

Worldwide Distribution - USA (nationwide) including the states of: CA, FL, GA, IL, KS, MA, MD, MO, NE, NY, and TX and the countries of: Canada, Germany, Romania, Spain, Palestine, South Africa, France, Sweden and Singapore

Lot / Code Info

Model #D3000I

Root Cause

Error in labeling

Action Taken

Linvatec Corp sent an "URGENT: Medical Device Recall Notification" letter dated January 24, 2013 to its customers after initial phone notifications. The letter describes the affected product, problem and actions to be taken. Customers were instructed to immediate check facility's inventory for the affected product, use the enclosed FedEx air-bill when returning the product, and complete and return to the Recall Reply Form. For questions contact your Customer Service department at 800-535-8536 or fax at 727-319-5701 or email Custserv1@linvatec.com.