CELL-DYN 3200 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.

Company
Abbott Laboratories
Recall Initiated
April 16, 2010
Posted
September 2, 2010
Terminated
December 6, 2010
Recall Number
Z-2350-2010
Quantity
69 units total.
Firm Location
Santa Clara, CA

Reason for Recall

Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.

Distribution

Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.

Lot / Code Info

Reservoir Date Codes Dec 17 2009 to March 17, 2010. Part Number/Serial numbers-- CS: 04H59-01/30664AH, 4H59-03/30276AH96, 30671AH96, 30673AH96. SL: 04H60-01, 62794AF-62799AF, 62144AF, 62772AF, 62073AF, 61628AF, 61376AF

Root Cause

Nonconforming Material/Component

Action Taken

Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.