Focal Radiation Treatment Planning System, Focal Release 4.3.1 through 4.34.02. For use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.

This recall has been terminated (originally issued June 15, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
March 31, 2008
Posted
June 15, 2011
Terminated
April 9, 2014
Recall Number
Z-2572-2011
Quantity
307
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

Focal: In absolute dose mode a beam with a port may be calculated with open normalization, with the result that the displayed dose reflects open normalization while the monitor units and QA plan reflect blocked normalization. This is particularly important when these beams also include compensators. The problem is detectable by looking at the delivered dose to the weight point in the patient as

Distribution

Worldwide Distribution -- USA (nationwide) and the countries of Albania, Argentina, Australia, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, France, Germany, Greece, India, Italy, Japan, Netherlands, Peru, Portugal, Qatar, Romania, Russia, Serbia, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, and Venezuela.

Lot / Code Info

Focal Release 4.3.1 through 4.34.02

Root Cause

Software design

Action Taken

Computerized Medical Systems Inc sent a "CUSTOMER ADVISORY" letter dated March 2008 to all affected customers. The letter describes the product, the problem and the actions to be taken. The firm provided a workaround for customers that states "Alternatively the plan can be imported directly into XIO". A copy of the Advisory was shipped with software within the affected range to those requesting a software upgrade. The issue was resolved in Focal Release 4.40.00 introduced in June 2008. The firm will send out a User Notice to all sites still running affected software to remind them that the issue resolution is available. For further questions, please call (403) 830-8023.