MRP-7000 and AIRIS Magnetic Resonance Imaging Systems, Software Versions: V7.0A to V7.0J.

This recall has been terminated (originally issued October 12, 2006).

Company
Hitachi Medical Systems America Inc
Recall Initiated
August 16, 2006
Posted
October 12, 2006
Terminated
April 13, 2012
Recall Number
Z-0029-2007
Quantity
348 units
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Image orientation error. When a 3D Maximum Intensity Projection (MIP) image data set is transferred from the MRI system to a computer workstation via the DICOM protocol, the anatomical markers do not change between the images as they rotate.

Distribution

Nationwide and Puerto Rico.

Lot / Code Info

AIRIS - Serial Number Range: A001, AG001-AG039, A901; MRP-7000 - Serial Number Range: 7001- 7154.

Root Cause

Other

Action Taken

On 8/16/2006, Hitachi issued two (2) Device Correction letters. The first letter was addressed to the five (5) customers whose system units are currently ''de-installed''. This letter informs the customers of the problem and asks that the customer(s) please contact the firm''s Customer Service Center on their toll-free tel. # 1-800-800-4925, in order to schedule a device correction should the customer ever re-activate/install their system(s). The second letter, issued to the customers with activated MRP-7000 and/or AIRIS MRI systems, informs the customers of the problem, and states that a HMSA Service (person) will be contacting the customer in order to schedule a device correction which should address the problem.