NeoPAP Neonatal CPAP/Humidification System (Ventilator, Continuous)

This recall has been terminated (originally issued April 7, 2006).

Company
Respironics, Inc.
Recall Initiated
December 27, 2005
Posted
April 7, 2006
Terminated
March 22, 2007
Recall Number
Z-0717-06
Quantity
83 effected units were identified
Firm Location
Murrysville, PA
Official Source
View on FDA website ↗

Reason for Recall

Vapotherm humidification cartridges were found to contain microbial (Ralstonia spp.) cultures. This device uses cartridges from the same supplier.

Distribution

The product was shipped to medical facilities in GA, IA, MA, MN, NJ, OK, PA, TX, and WI. The product was also shipped to Sweden, South Africia, Egypt, Greece, and Saudi Arbia.

Lot / Code Info

Lot codes: 3124050609-01, 3124050609-02, 3124050609-03, 3124050803-01, 3124050803-02, 3124050803-03, 3124050803-04, 3124050803-05, 3124050623-01, 5313050323-01, 5313050620-01, 5306050624-01, 5306050623-01, 5306050323-01, 4568050622-01, 4568050415-01, 3195050217-06, 3195050511-01, 3195050426-12, 3195050511-02, 3195050511-03, 3195050511-05, 3195050511-07, 3195050511-08, 3195050511-10, 3195050520-07, 3195050412-05, 3195050418-02, 3195050418-03, 3195050418-06, 3195050418-07, 3195050418-09, 3195050418-13, 3195050418-14, 3195050418-15, 3195050426-01, 3195050426-02, 3195050426-03, 3195050426-04, 3195050426-05, 3195050426-11, 3195050426-13, 3195050426-14, 3195050426-15, 3195050511-09, 3195050511-11, 3195050511-12, 3195050511-13, 3195050511-14, 3195050511-15, 3195050520-01, 3195050520-02, 3195050520-03, 3195050520-04, 3195050520-05, 3195050527-02, 3195050527-03, 3195050527-04, 3195050527-05, 3195050527-06, 3195050527-07, 3195050527-08, 3195050131-08, 3195050120-02, 3195050120-07, 3195050120-08, 3195050120-09, 3195050114-03, 3195050527-12, 3195050527-13, 3195050527-14, 3195050527-15, 3195050609-01, 3195050609-02, 3195050609-04, 3195050722-07, 3195050609-07, 3155050722-04, 3155050722-05, 3195050811-03, 3195050426-07, 3195050609-09, 3195050811-04

Root Cause

Other

Action Taken

Respironics contacted customers via tracked letter informing them to discontinue use of their NeoPAP system. Distributors were contacted via telephone and e-mail instructing them to contact their customers and instruct them to discontinue use of the NeoPAP system. The recalling firm issued a Customer Advisory Notice (CAN) dated 1/5/06 to all customers.