RecallDepth

Medline RF Detect Sterile X-Ray Detectable USP Type VII Gauze; 8" x 4", 12-ply, 100% cotton sponges, 10 sponges per pack, 80 packs per case; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA, Made in China; reorder: NON21432RF

Company
Medline Industries Inc
Recall Initiated
June 13, 2008
Posted
September 20, 2008
Terminated
January 9, 2009
Recall Number
Z-2321-2008
Quantity
295 cases
Firm Location
Mundelein, IL

Reason for Recall

There is the potential for the RF ID tag to separate from the retaining pouch.

Distribution

California, Indiana, Missouri, Ohio, Oklahoma, Pennsylvania, Texas and Washington

Lot / Code Info

reorder: NON21432RF, lots CM130002052, CM130002062, CM130002075, CM130002068, CM130002079, and CM130002110

Root Cause

Device Design

Action Taken

Initially a recall letter dated 6/13/08 was sent to the direct accounts identifying 5 lots of affected lots with two reorder numbers involved, explaining that there was a lack of integrity of the retaining pouch material for the RFID gauze, and that there was a potential for the RFID tag to separate from the retaining pouch during use. It was subsequently discovered that there was only one reorder number affected by the problem and that an additional lot had been gamma sterilized. Therefore a second letter dated 6/19/08, listing the correct reorder number and the six affected lots, was sent via FedEx overnight on the same date. The accounts were requested to remove RFID gauzes with the listed lot numbers from all stock rooms, materials storage, and OR's, and return it to Medline for replacement. The accounts were requested to complete the enclosed response sheet and fax it back to Medline at 847-643-4482, indicating the amount and lot numbers of the gauzes being returned. Any questions were directed to 866-359-1704, between 8:00 AM and 5:00 PM CST. Distributors were requested to notify their accounts of the recall. If you have questions about this recall, contact Kathy Dunne at 847-970-4678 or at kdunne@medline.com.

More recalls by Medline Industries Inc