RecallDepth

Polaris Dx Diagnostic Catheter, for cardiac electrophysiology studies, Model M0049663S0

This recall has been terminated (originally issued August 22, 2007).

Company
Boston Scientific Corp
Recall Initiated
June 26, 2007
Posted
August 22, 2007
Terminated
November 27, 2007
Recall Number
Z-1191-2007
Quantity
30 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Product is mislabeled as hexapolar (6 electrodes), but is actually quadripolar (4 electrodes).

Distribution

Product was distributed to 3 consignees in Florida and Tennessee.

Lot / Code Info

Batch numbers 9535942, 9596195, 9618290. Catalog Model number/Product UPN number M00496630

Root Cause

Labeling mix-ups

Action Taken

The firm issued notification letters to its consignees, sent 06/26/2007, via FedEx. The firm intends to remove the product and destroy it.

More recalls by Boston Scientific Corp

View all recalls by this company →