IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-8 Material Number: M004PM4500N40; Catalog number: PM4500N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia typically chronic, drug refractory atrial fibrillation.

This recall has been terminated (originally issued November 20, 2015).

Company
Boston Scientific Corp
Recall Initiated
November 20, 2015
Terminated
November 27, 2019
Recall Number
Z-0464-2016
Quantity
1556 total - all models
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSCs product risk documentation.

Distribution

Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Netherlands, Saudi Arabia, Singapore, Thailand, and South Africa.

Lot / Code Info

Material Number: M004PM4500N40;  Catalog number: PM4500N4; Lot numbers: 16309584, 16615976, 16623754, 16623755, 16722434, 16736927, 16739588, 16743271, 16872124, 16999075, 17118111, 17157418, 17201805, 17222051, 17222052, 17370483, 17376573, 17376574, 17436562, 17436563, 17436565, 17451802, 17478950, 17499145, 17499152, 17590503, 17611785, 17625992, 17768306, 17791420, 17925492, 17925493, 18012208, 18161715. Expiration Date Range:8/15/2016 to 7/2/2018.

Root Cause

Device Design

Action Taken

Boston Scientific sent an Urgent Medical Device Removal letters to all affected customers dated November 20, 2015 via Fed Ex or other equivalent service. The letter identified the affected products, problem and actions to be taken. Customers were instructed to contact their local Sales Representative to answer any questions.