IntellaTip MiFi Open-Irrigated radiofrequency Ablation Catheters: Material number: M004EPM96200, Catalog Number: EPM9620: Material number: M004EPM9620K20, Catalog Number: EPM9620K2; Material number : M004EPM9620N40; Catalog number: EPM9620N4. The IntellaTip MiFi Open-Irrigated Ablation Catheter is a 7.5F (2.5 mm) quadrapolar open-irrigated (OI) ablation catheter designed to deliver radiofrequency (RF) energy to the 4.5 mm catheter tip electrode for cardiac ablation The product is not approved for use/sale in the US.

This recall has been terminated (originally issued March 2, 2015).

Company
Boston Scientific Corp
Recall Initiated
February 3, 2015
Posted
March 2, 2015
Terminated
March 21, 2017
Recall Number
Z-1209-2015
Quantity
14 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Reports of formation of char adherent to the proximal part of the distal tip electrodes

Distribution

Distributed in Germany, Great Britain, and the Netherlands.

Lot / Code Info

Material number: M004EPM96200, Catalog Number EPM9620; Serial numbers: 16941576, 16943730, 17150129, 17156568, 17156570, 17168752, 17171230, 17218387, 17221754, 17237747, 17247469, 17261879.  Material number: M004EPM9620K20, Catalog Number: EPM9620K2: Serial numbers : 17018928, 17018929, 17147617, 17172602, 17175351, 17223957, 17230227, 17233882, 17235011, 17241926, 17263604, 17285884, 17288714.  Material number : M004EPM9620N40; Catalog number: EPM9620N4; Serial numbers: 17148152, 17226126, 17266642.  All with Expiry Dates of April 20, 2015 - June 14, 2015.

Root Cause

Device Design

Action Taken

Customer notification letters were sent by overnight mail on February 10, 2015.