Medline Industries, Inc. "Cedar" brand Vinyl Economy Exam Glove, Powder-Free, Large Size, Catalog Number: MSV403P, Lot Number: JCU04958569
- Company
- Medline Industries Inc
- Recall Initiated
- February 3, 2011
- Posted
- March 4, 2011
- Terminated
- March 19, 2012
- Recall Number
- Z-1554-2011
- Quantity
- 696 cases (696,000 units)
- Firm Location
- Mundelein, IL
Reason for Recall
The recall was initiated because the affected lot, lot number JCU004958569, of gloves did not meet the requirements for the water-leak test.
Distribution
Nationwide Distribution including CA, HI, and NV.
Lot / Code Info
Lot Number: JCU04958569
Root Cause
Release of Material/Component prior to receiving test results
Action Taken
MEDLINE INDUSTRIES, INC. sent a RECALL NOTICE IMMEDIATE ACTION REQUIRED letter by US Mail on February 10, 2011, to all affected cusomters. The letter identified the product, the problem, and the action to be taken by the customer. The customers were instructed to: 1. Immediately quarantine ALL inventory of the affected lot for return to Medline. 2. Complete the attached recall verification form. Please complete this form EVEN IF you have none of the affected lot in stock. 3. Fax or email the completed form to Medline. The email address is Recalls@Medline.com or faxed the form to 866-767-1290. Customers with questions regarding the recall were instructed to contact Medline at 1-866-359-1704 or via e-mail at Recalls@Medline.com.