XiO Radiation Treatment Planning System, XiO Release 1.5.0 and above. Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.

This recall has been terminated (originally issued June 15, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
May 14, 2008
Posted
June 15, 2011
Terminated
April 8, 2014
Recall Number
Z-2574-2011
Quantity
936
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

In XiO, when Telepathy is exited, re-entered, and the temporary plan re-called, the OLD beam weight will be displayed but the dose distribution will reflect the modified beam weight.

Distribution

Worldwide Distribution -- USA (nationwide) including Puerto Rico, and countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece, Guatemala, Hungary, India, Indonesia, Israel, Italy, Japan, Libya, Lithuania, Malaysia, Mexico, Morocco, .Netherlands, New Zealand, Nicaragua, North Korea, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Turkey, Ukraine, United Kingdom, Uruguay, Venezuela, Yemen, and Zimbabwe.

Lot / Code Info

XiO Release 1.5.0 and above

Root Cause

Software design

Action Taken

Computerized Medical Systems, Inc. sent an CUSTOMER ADVISORY letter to all affected customers in May 2008. The letter identified the product, the problem and the action needed to be taken by the customer. The issue was resolved in XiO Release 4.50.00 which has been available since June 2009. A User Notice was sent in March 2011, to all the customers reminding them of the availability of the correction. This notice was sent hardcopy via US Mail but there was no return receipt. For further questions, please call ( 408) 380-8023