RecallDepth

MicroAire PAL Cannula Product Usage - Power Aspiration Device

This recall has been terminated (originally issued February 8, 2016).

Company
Microaire Surgical Instruments, Llc
Recall Initiated
February 8, 2016
Terminated
February 14, 2017
Recall Number
Z-1328-2016
Quantity
145 manufactured and distributed
Firm Location
Charlottesville, VA
Official Source
View on FDA website ↗

Reason for Recall

MicroAire has initiated a recall on product PAL-R4011XL (40cm x 4mm Del Vecchio French Rasp cannula), due to our receipt of three (3) complaints that it bent or broke during surgery. Accordingly, the PAL-R4011XL has also been put on production hold until further notice.

Distribution

Worldwide Distribution - US Nationwide

Lot / Code Info

Include lot numbers/serial numbers  0913228216, 0414243005, 0215266350, 0415269875, 0515271334

Root Cause

Nonconforming Material/Component

Action Taken

The firm is sending letters to all customers who have received PAL-R4011XL from August 2013- October 2015. The firm sent these letters by UPS to expedite the delivery and for traceability.